Studies of slip gauge calibration using TESA UPC system / Ahmedsyah Abdullah, Rosnadiah Bahsan and Norlela Wati Abu Bakar

Abdullah, Ahmedsyah and Bahsan, Rosnadiah and Abu Bakar, Norlela Wati (2001) Studies of slip gauge calibration using TESA UPC system / Ahmedsyah Abdullah, Rosnadiah Bahsan and Norlela Wati Abu Bakar. [Student Project] (Unpublished)

Abstract

This project has been carried out to study how slip gauge is being calibrated using TESA UPC Calibration System. Calibration process of slip gauge was investigated with master gauge blocks served as reference to the gauge blocks to be calibrated using TESA UPC Calibration System. The main objective was to determine the grade of steel gauge blocks either as confirmed or a new grade is given. Before the TESA UPC Calibration System is automatically processes and certifies measured values, ensured the temperature between measuring stand table, master and test gauge block to be calibrated at stable temperature 20 + 0.5°C according to the coefficient of expansion. Actually, the ensure the temperature to be stable, the measurement should only begin after wiped the gauge blocks with a cotton swab dipped in vesseline. Whereas in this project, the gauge blocks is not followed the process. It is because in the laboratory has not the kit to wipe the gauge blocks. After wiped the gauge blocks, left it for about 24 hours (1 day) to stabilize with the surrounding temperature and then its can be measured. The greater the different from the reference temperature of 20 ± 0.5°C, the greater the measuring uncertainty will be. The exact result dependent to the temperature stabilization. Process calibration of steel gauge blocks grade 1 using comparative measurement method of TESA UPC Calibration System was determined with ensured the temperature reference is stabilized at 20 + 0.5°C was achieved and the wiped of gauge blocks in the right method.

Metadata

Item Type: Student Project
Creators:
Creators
Email / ID Num.
Abdullah, Ahmedsyah
98079032
Bahsan, Rosnadiah
98079608
Abu Bakar, Norlela Wati
98079542
Contributors:
Contribution
Name
Email / ID Num.
Thesis advisor
Kiyai Abas, Sunhaji
UNSPECIFIED
Subjects: T Technology > T Technology (General) > Industrial research. Research and development
T Technology > TA Engineering. Civil engineering > Surveying
Divisions: Universiti Teknologi MARA, Shah Alam > Faculty of Mechanical Engineering
Programme: Diploma in Mechanical Engineering (Manufacturing)
Keywords: Calibration, slip gauge, accuracy of measurement
Date: 2001
URI: https://ir.uitm.edu.my/id/eprint/73151
Edit Item
Edit Item

Download

[thumbnail of 73151.pdf] Text
73151.pdf

Download (122kB)

Digital Copy

Digital (fulltext) is available at:

Physical Copy

Physical status and holdings:
Item Status:
On Shelf

ID Number

73151

Indexing

Statistic

Statistic details