Abstract
This project has been carried out to study how slip gauge is being calibrated using TESA UPC Calibration System. Calibration process of slip gauge was investigated with master gauge blocks served as reference to the gauge blocks to be calibrated using TESA UPC Calibration System. The main objective was to determine the grade of steel gauge blocks either as confirmed or a new grade is given. Before the TESA UPC Calibration System is automatically processes and certifies measured values, ensured the temperature between measuring stand table, master and test gauge block to be calibrated at stable temperature 20 + 0.5°C according to the coefficient of expansion. Actually, the ensure the temperature to be stable, the measurement should only begin after wiped the gauge blocks with a cotton swab dipped in vesseline. Whereas in this project, the gauge blocks is not followed the process. It is because in the laboratory has not the kit to wipe the gauge blocks. After wiped the gauge blocks, left it for about 24 hours (1 day) to stabilize with the surrounding temperature and then its can be measured. The greater the different from the reference temperature of 20 ± 0.5°C, the greater the measuring uncertainty will be. The exact result dependent to the temperature stabilization. Process calibration of steel gauge blocks grade 1 using comparative measurement method of TESA UPC Calibration System was determined with ensured the temperature reference is stabilized at 20 + 0.5°C was achieved and the wiped of gauge blocks in the right method.
Metadata
Item Type: | Student Project |
---|---|
Creators: | Creators Email / ID Num. Abdullah, Ahmedsyah 98079032 Bahsan, Rosnadiah 98079608 Abu Bakar, Norlela Wati 98079542 |
Contributors: | Contribution Name Email / ID Num. Thesis advisor Kiyai Abas, Sunhaji UNSPECIFIED |
Subjects: | T Technology > T Technology (General) > Industrial research. Research and development T Technology > TA Engineering. Civil engineering > Surveying |
Divisions: | Universiti Teknologi MARA, Shah Alam > Faculty of Mechanical Engineering |
Programme: | Diploma in Mechanical Engineering (Manufacturing) |
Keywords: | Calibration, slip gauge, accuracy of measurement |
Date: | 2001 |
URI: | https://ir.uitm.edu.my/id/eprint/73151 |
Download
73151.pdf
Download (122kB)