Determination of phenol in wastewater sample by using high performance liquid chromatography (HPLC) / Nurul Kudsiah Hj Katemon

Katemon, Nurul Kudsiah (2017) Determination of phenol in wastewater sample by using high performance liquid chromatography (HPLC) / Nurul Kudsiah Hj Katemon. [Student Project] (Unpublished)
Abstract

The presence of phenol in water is risky to living organism as the toxicity effect causes the long-term exposure. In this research, the analysis was accomplished on a Cl8 column with UV-Vis detector (HPLC-UV) for determination of phenol in wastewater water while the solid phase extraction (SPE) is used as sample preparation technique. Five mL of load sample was used as the optimized variable in SPE while 0.75 mL/min was used as flow-rate on eluent in HPLC. The analytical method was validated according to the following parameter, which are precision, linear range, limit of detection (LOD), and limit of quantitation (LOQ). A good linear correlation coefficient with R2 = 0.9991 was observed over the range of 3.05 to 30.0 μg/mL. The limit of detection was calculated to be 0.41 μg/mL, while the limit of quantitation value of the validated method was measured to be 1.37 μg/mL. Good recoveries were obtained in the range between 90 to 120%. In conclusion, the proposed method was found to be appropriate and accurate for the determination of phenol in wastewater by using HPLC.

Item Details
Edit Item
Edit Item
Downloads & Files
[thumbnail of PPb_NURUL KUDSIAH HJ KATEMON AS N 17_5.pdf]
Text
PPb_NURUL KUDSIAH HJ KATEMON AS N 17_5.pdf
Download (103kB)
Location & Physical Holdings
Digital Copy
Physical Location
Item Status
Indexing & Metrics